Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this icd were reviewed.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.There was no indication of a material or manufacturing problem.Subsequently the device was interrogated.The interrogation could be properly performed and revealed the eos battery status, detected on january 21, 2026.All electrical parameters were verified and showed a normal device behavior.Especially, the current consumption and the battery state of discharge were found normal and as expected.The memory content of the device was inspected.During the analysis of the available iegms noise was observed in the right ventricular channel, leading to multiple charging cycles that resulted in several shock deliveries, confirming the clinical observation.However, there was no indication of a device malfunction.
|