• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LUTONIX, INC LUTONIX 018 DRUG COATED DILATATION CATHETER; DRUG COATED BALLOON PTA CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LUTONIX, INC LUTONIX 018 DRUG COATED DILATATION CATHETER; DRUG COATED BALLOON PTA CATHETER Back to Search Results
Model Number 9014
Medical Device Problem Codes Inflation Problem (1310); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/13/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: as the lot number for the device was provided, a review of the device history records was performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.Section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
On (b)(6) 2026, a patient underwent an angioplasty procedure by using the lutonix 018 drug coated balloon.It was reported that during the procedure, the balloon allegedly twisted during inflation.The procedure was completed using the same device.There was no reported patient injury.
 
Additional Manufacturer Narrative
H11: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the reported material twist during inflation could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.B5, d4 (medical device model #), g3, h6 (method).Section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
On (b)(6) 2026, a patient underwent an angioplasty procedure by using the lutonix 018 drug coated balloon.During the procedure, it was reported that the balloon allegedly twisted during inflation.The procedure was completed using the same device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUTONIX 018 DRUG COATED DILATATION CATHETER
Common Device Name
DRUG COATED BALLOON PTA CATHETER
Manufacturer (Section D)
LUTONIX, INC
9409 science center dr
new hope MN 55428
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key24265380
Report Number3006513822-2026-00004
Device Sequence Number10031177
Product Code ONU
UDI-Device Identifier00801741184727
UDI-Public(01)00801741184727(11)250902(17)280820(10)GFKT2692
Combination Product (Y/N)Y
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P130024
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9014
Device Catalogue NumberLX1813063005F
Device Lot NumberGFKT2692
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/13/2026
Supplement Date Received by Manufacturer02/13/2026
Initial Report FDA Received Date02/05/2026
Supplement Report FDA Received Date03/02/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/02/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-