| Model Number |
1899200 |
| Medical Device Problem Codes |
Overheating of Device (1437); Power Problem (3010)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/30/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: product analysis unable to perform temperature test as motor was not rotating.Found the handpiece cosmetically not ok.Connector has dent.Thumb wheel jammed and damaged.Hi-pot test fail.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that preoperative, handpiece was set with 25cc/min irrigation, 3000 rpm in oscillation mode.Handpiece was stuck and overheating within a minute.There was no patient involved.
|
| |
|