RESPIRONICS, INC. DREAMSTATION BIPAP AVAPS25; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING Back to Search Results
Model Number INX1125S19
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/30/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The capa process identified complaints not being created for quality issues found during third party servicing.Based on a review of records impacted by the identified issue, it was determined that this record needed to be corrected and is now being submitted accordingly.
 
Event or Problem Description
A dreamstation bipap avaps25 device was returned to a service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.During the evaluation of the device at the service center, the device was visually inspected and visible foam particles were observed.Additionally, the patient¿s complaint was confirmed, and the device was scrapped.At this time, no further investigation can be performed.If any additional information is received, a follow up report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION BIPAP AVAPS25
Common Device Name
VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key24267115
Report Number2518422-2026-100489
Device Sequence Number10723436
Product Code MNS
UDI-Device Identifier00606959425040
UDI-Public(01)00606959425040
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K102465
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 02/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberINX1125S19
Device Catalogue NumberINX1125S19
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/30/2025
Initial Report FDA Received Date02/05/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/11/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1973-2021
Patient Sequence Number1
Patient SexUnknown
-
-