| Catalog Number |
950014 |
| Medical Device Problem Code |
Loss of or Failure to Bond (1068)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/23/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a pfc sig rpf cem fem sz 3 lt, unknown knee tibial tray, unknown knee tibial insert, and unknown knee patella were involved in a reported event.The event concerned revision surgery of a knee construct, including a femoral component, tibial tray, tibial insert, and patella.The revision was necessitated by loosening attributed to poly lysis, with medial wear observed on the insert.Both the femur and tibia were found to be loose, while the patella was revised due to polyethylene wear.All components were explanted.The patient required a corrective invasive procedure in which the devices were replaced or revised.The loosening was identified at the implant to cement interface, and wear or deterioration of polyethylene material was noted.Osteolysis was also referenced as a related condition.Implanted in 2006.Explanted (b)(6) 2026.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary- no device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed as the required product/lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot- the device lot number is unknown, therefore a¿device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.Added: d10 concomitant.
|
| |
|
Search Alerts/Recalls
|
|