Additional information was provided in sections d9, h3, h.6 and h.11.The returned instrument was received at the investigation site in its inner and outer blister, covered with the tyvek foil shows the lot information.However, the instrument was not correctly insert at the inner blister.The product shows macroscopic signs of damage, namely the forceps arms are bent.There are no signs of surgery residues.The returned instrument was visually inspected with the aid of a photomicroscope using various magnifications.The visual inspection confirmed the macroscopically noticeable damage and shows the bent forceps arms in detail.However, the point in time when the described damage occurred can no longer be determined conclusively nor can the situation that led to it be reconstructed.The customer¿s complaint is confirmed as the returned device shows deformation, but a root cause for the reported issue cannot be determined.A review of the device history record traceable to the reported lot numbers, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.Non-conformance report provided by originator.A non-conformance-based review of the lot number was performed and did not reveal any potential contributing factors to the reported complaint.It cannot be determined how or where the damage to the instrument occurred.Therefore, the root cause for the customer's reported event could not be determined.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all manufacturer products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.The complaint data will continue to be monitored for evidence of adverse trending and further action will be taken, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
|