Model Number SA60AT
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/09/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information was provided in d.10., h.3., h.6, and h.11 the product was not returned.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.The account indicated the use of a qualified associated cartridge and handpiece with a non-qualified viscoelastic.The root cause for the reported complaint may be related to a failure to follow the instructions for use (ifu).The viscoelastic indicated is not qualified for the lens/company cartridge combination used.The ifu instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ophthalmic visco surgical device (ovd)-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A -healthcare professional reported that before intraocular implantation surgery it was noticed that haptic was bent.Surgery was completed on the same day.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key24267881
Report Number1119421-2026-00237
Device Sequence Number21511394
Product Code HQL
UDI-Device Identifier00380652555784
UDI-Public(01)00380652555784(17)290928(11)240930(10)15928722
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSA60AT
Device Catalogue NumberSA60AT.190
Device Lot Number15928722
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/12/2026
Supplement Date Received by Manufacturer02/27/2026
06/19/2026
Initial Report FDA Received Date02/05/2026
Supplement Report FDA Received Date03/18/2026
07/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/30/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HYALURON; MONARCH III IOL, CARTRIDGE C; MONARCH III IOL, INJECTOR
Patient SexUnknown
-
-