This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation code was added: 10, 4109, 3331 and 4111.H6: investigation findings code was added: 213.H6: investigation conclusions code was added: 4315 and 67.H10: narrative/data was updated.Zimvie received the reported device for evaluation.Visual evaluation was performed, abutment has no damage.Functionally tested in-house implant and engages, retains/seats and disengages as intended.No malfunction identified.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.The customer did not submit images for the reported event.A definitive root cause could not be determined since device malfunction did not occur.The reported event was unconfirmed following functional testing and physical evaluation.Therefore, based on the available information, a device malfunction did not occur.No malfunction identified during functional testing.The reported event was unconfirmed.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
|