| Catalog Number |
00633405028 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Bone Fracture(s) (1870); Joint Dislocation (2374); Osteolysis (2377)
|
| Date of Event |
12/23/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: s&n synergy stem.G2: foreign - event occurred in canada.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient experienced a closed reduction and is pending a left hip revision with a triflange due to dislocation, osteolysis, and non-displaced superior and inferior pubic rami fractures.The patient has some instability and the locking mechanism for components apparently failed.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.\
\
the following sections were updated: d4; d10; h2; h3; h4; h6; h10.\
\
d10: s&n synergy stem and s&n head.\
\
complaint is not confirmed.\
\
no product was returned or pictures provided; visual and dimensional evaluations could not be performed.\
\
review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.\
\
review of complaint history identified additional similar complaints for the reported item and no additional complaints for the reported part and lot combination.\
\
medical records were provided and reviewed.Review identified the following: limited information provided.No part/lot information or official operative records were provided, only progress notes and radiology reports.Emails indicate zb implants are involved and a pmi request for a triflange component.Primary left tha on an unknown date with a history of multiple revisions due to aseptic loosening.No information was provided to indicate what components were revised or when.Left hip dislocated and underwent closed reduction.Severe osteolysis and non-displaced superior and inferior pubic rami fractures were noted.\
\
a definitive root cause cannot be determined; however, there was a competitor stem and head used with a zimmer biomet head.It is unknown if this caused or contributed to the reported event.\
\
if any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further information is available at the time of this report.
|
| |
|