| Catalog Number |
165808 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/22/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that at the end of the case, a foley catheter was placed.During balloon inflation, resistance was noted at about 2ml, so inflation was stopped.The balloon would not deflate with aspiration, so the inflation port was cut, allowing the balloon to deflate.The foley was then removed easily without issue.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation.The reported event is addressed within the labeling as it states: instructions for use: follow your facility protocol for all processes related to catheterisation, re catheterisation, stabilization and urine collection.Follow aseptic procedures outlined by your local hospital or healthcare facility.Thoroughly wash hands, wear sterile gloves, and use aseptic technique whenever handling the catheter or catheter components during insertion.Visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.No additional actions can be taken at this time.Corrections: d, f, h.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.Udi format updated with available product information per gudid.H11: sections a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that at the end of the case, a foley catheter was placed.During balloon inflation, resistance was noted at about 2ml, so inflation was stopped.The balloon would not deflate with aspiration, so the inflation port was cut, allowing the balloon to deflate.The foley was then removed easily without issue.
|
| |
|