|
It was reported that unstable speed occurred.The target lesion was located in the coronary artery.A rotapro 1.75mm was selected for use.During rotation in the high-speed mode, it was noted that although the set rotational speed was 140,000, the maximum rotational speed was found to be 160,000 rpm.Additionally, it was noted that during rotational grinding a stall occurred.The procedure was completed with another of the same device.There were no patient complications, and the patient was stable post procedure.
|