• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON VISCOUS FLUID CONTROL TUBING SET (SILICONE OIL INJ/EXT); TUBING, NONINVASIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON VISCOUS FLUID CONTROL TUBING SET (SILICONE OIL INJ/EXT); TUBING, NONINVASIVE Back to Search Results
Catalog Number 8065750957
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/09/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non-healthcare professional reported that there was an issue occurred at the end of the silicone oil tubing, could not inject silicone oil properly during vitrectomy surgery.The surgery was completed on the same day after replacing the product with another one.There was no impact to the patient.
 
Additional Manufacturer Narrative
Additional information provided in d.9., h.3., h.6.And h.11.The returned product was visually inspected and no obvious defects were observed that would have contributed to the reported event.Additionally, all connections were found to fit securely.A calibrated console was used to test the sample.The consoles could recognize the viscous fluid control (vfc) by illuminating green on the light emitting diode (led) ring.The vfc sample was filled with silicone oil per directions for use (dfu); in injection mode the plunger moved freely and met specifications.Extraction mode was then selected and the vfc could aspirate silicone oil into the barrel of the syringe; no anomalies were observed during this step.Pressure was stable during functional testing.The plunger in the syringe moved smoothly during operation.The investigation conducted a non-conformance review of the reported lot number.No deviation were identified that would have contributed to this event and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event as the product functioned per specifications.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.Based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint, therefore no action will be taken for this occurrence.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISCOUS FLUID CONTROL TUBING SET (SILICONE OIL INJ/EXT)
Common Device Name
TUBING, NONINVASIVE
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key24269099
Report Number1644019-2026-00688
Device Sequence Number15939666
Product Code GAZ
UDI-Device Identifier00380657509577
UDI-Public(01)00380657509577(17)270228(11)250311(10)17A0P6
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065750957
Device Lot Number17A0P6
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/14/2026
Supplement Date Received by Manufacturer06/11/2026
Initial Report FDA Received Date02/05/2026
Supplement Report FDA Received Date07/01/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/11/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-