Model Number UL-878
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/19/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a procedure, upon opening the packaging, broken threads were found in six packs.To resolve the issue, a new device was used.No consequences or impact to the patient were reported.
 
Additional Manufacturer Narrative
D10.Concomitant products: ul-878, ul-878 polysorb* 2-0 vio 75cm gu46 x36 (lot d5h3844y); ul-878, ul-878 polysorb* 2-0 vio 75cm gu46 x36 (lot d5h3844y); ul-878, ul-878 polysorb* 2-0 vio 75cm gu46 x36 (lot d5h3844y); ul-878, ul-878 polysorb* 2-0 vio 75cm gu46 x36 (lot d5h3844y); ul-878, ul-878 polysorb* 2-0 vio 75cm gu46 x36 (lot d5h3844y).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6 correction: e1(street 1 and postal code) h3 evaluation summary: medtronic conducted an investigation based upon all information received.One representative sample was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that there were broken threads.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYSORB
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC AC
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key24271887
Report Number9612501-2026-00288
Device Sequence Number18822193
Product Code GAM
UDI-Device Identifier10884521049635
UDI-Public(01)10884521049635(17)300731(10)D5H3844Y(20)01
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K963253
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberUL-878
Device Catalogue NumberUL-878
Device Lot NumberD5H3844Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/04/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/20/2026
Supplement Date Received by Manufacturer04/07/2026
Initial Report FDA Received Date02/05/2026
Supplement Report FDA Received Date04/15/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/25/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Patient SexUnknown
-
-