• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 20 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 20 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number 20-600
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 01/19/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported "ruptured implant".Later healthcare professional reported "total rupture".Patient underwent capsulectomy.This record is for the left side.Device was explanted.
 
Additional Manufacturer Narrative
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
 
Additional Manufacturer Narrative
Device evaluation: per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: ¿ rupture: observed broken device through microscopic inspection assessed as surgical damage and missing piece of shell assessed as inconclusive.No further actions are required as it is not related to the manufacturing process.No additional observations performed on the device.No further actions are required.
 
Event or Problem Description
Healthcare professional reported "ruptured implant".Later healthcare professional reported "total rupture".Patient underwent capsulectomy.This record is for the left side.Device was explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 20 SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
SANTA BARBARA
900 parkway global park zona f
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
8479366324
MDR Report Key24272741
Report Number9617229-2026-02426
Device Sequence Number9948233
Product Code FTR
UDI-Device Identifier10888628001763
UDI-Public10888628001763
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2005
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number20-600
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/19/2026
Supplement Date Received by Manufacturer04/22/2026
Initial Report FDA Received Date02/05/2026
Supplement Report FDA Received Date05/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age52 YR
Patient SexFemale
Patient Weight59 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-