Catalog Number 242137
Medical Device Problem Codes Break (1069); Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that prior to an unspecified surgical procedure it was observed that the 4k c-mnt scp, 4.0, 30, 167, mitek scope device was cracked.During in-house engineering evaluation it was determined that the device was broken (2+ pieces: chipped/shaved/delaminated).This event did not occur during a surgical procedure.There were no reported adverse consequences to the patient.No additional information could be provided.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: investigational summary: the complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: other damages caused by customer outer tube damaged, needle scratched outer tube damaged, distal tip distal tip damaged dings/ scratched/nicks illumination distal tip fiber damaged optical system, optical components broken lenses in optical system distal cover glass/negative damaged chipped cosmetic issue scratches/dents engraving/identification datamatrix code level a labeling: illegible etch, visual: deformed/bent, broken (2+ pieces): chipped/shaved/delaminated, visual: scratched/nicked.Per service reports, this complaint was confirmed.The defective parts need to be replaced to resolve the issues.Based on the investigation findings, the potential root cause is traced to faulty parts.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: upon further investigation it was determined that the device evaluation required an update.Investigation summary the complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: repair level 3 major repair - other damages caused by customer - outer tube damaged, needle scratched - outer tube damaged, distal tip distal tip damaged dings/ scratched/nicks - illumination distal tip fiber damaged - optical system, optical components broken lenses in optical system distal cover glass/negative damaged chipped - cosmetic issue scratches/dents - engraving/identification datamatrix code level a labeling : illegible etch, visual : deformed/bent, broken (2+ pieces) : chipped/shaved/delaminated, visual : scratched/nicked per service report, this complaint was confirmed.The housing, optical, tube and label were replaced to resolve the issues.After repair, the device was found to be working according to the specifications.Based on the investigation findings, the potential root cause is traced to faulty parts.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4K EPSCP SCP,4.0,30,167,MITEK
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
elba bello
325 paramount drive
raynham, MA 02767
9083863534
MDR Report Key24273396
Report Number1221934-2026-00593
Device Sequence Number10725864
Product Code HRX
UDI-Device Identifier10886705031993
UDI-Public(01)10886705031993
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K080560
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number242137
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/31/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/04/2026
Supplement Date Received by Manufacturer02/12/2026
Initial Report FDA Received Date02/05/2026
Supplement Report FDA Received Date02/12/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-