|
MEDTRONIC MEXICO S. DE R.L. DE CV CUSTOM TUBING PACK; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIO
|
Back to Search Results |
|
| Model Number |
BB8G43R7 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that prior to use of a custom tubing pack, it was reported that the black one way valve on the red suction line was cracked, and not functioning.A complete break was observed and that the device was replaced.There was no patient involvement, so no adverse effect occurred.
|
| |
|
Search Alerts/Recalls
|
|
|