|
RESPIRONICS, INC. DREAMSTATION BIPAP AVAPS25; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
|
Back to Search Results |
|
| Model Number |
INX1125S19 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/30/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The capa process identified complaints not being created for quality issues found during third party servicing.Based on a review of records impacted by the identified issue, it was determined that this record needed to be corrected and is now being submitted accordingly.
|
| |
|
Event or Problem Description
|
|
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.There was no report of patient harm or injury.The device was returned to third party service center.During the evaluation, the device was visually inspected and found presence of visible foam particles.The device was scrapped.The manufacturer is submitting a final report at this time.If pertinent information becomes available to the manufacturer at a later date, an addendum to this final report will be filed.
|
| |
|
Search Alerts/Recalls
|
|
|