• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH GAMMA4 - LAG SCREW D10.5X90MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH GAMMA4 - LAG SCREW D10.5X90MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 81600090S
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/08/2026
Type of Reportable Event Malfunction
Event or Problem Description
As reported: "during the gamma4 long nail technique, after inserting the nail, when attempting to insert the lag screw, the precision pin was inserted but could not be positioned correctly.Despite carefully drilling for the lag screw without a guide pin, it was impossible to drill to the appropriate depth beyond the lag screw hole.As the drill simply would not pass through, the precise length of the lag screw could not be determined.An approximate length was estimated and the lag screw inserted, but it too stopped midway.After switching to a nail with a different neck angle and reinserting it, we were able to place it in the optimal position.Postoperatively, calibration using the targeting device, nail, and lag screw employed for insertion revealed no issues.Consequently, the surgeon concluded that the device had deflected due to the patient's bone quality and shape, preventing insertion.".
 
Additional Manufacturer Narrative
Additional information has been requested and if received, will be provided in the supplemental report.
 
Additional Manufacturer Narrative
The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.However the event description clearly says "postoperatively, calibration using the targeting device, nail, and lag screw employed for insertion revealed no issues." if the device is returned or any further information is provided, the investigation report will be reassessed.
 
Event or Problem Description
As reported: "during the gamma4 long nail technique, after inserting the nail, when attempting to insert the lag screw, the precision pin was inserted but could not be positioned correctly.Despite carefully drilling for the lag screw without a guide pin, it was impossible to drill to the appropriate depth beyond the lag screw hole.As the drill simply would not pass through, the precise length of the lag screw could not be determined.An approximate length was estimated and the lag screw inserted, but it too stopped midway.After switching to a nail with a different neck angle and reinserting it, we were able to place it in the optimal position.Postoperatively, calibration using the targeting device, nail, and lag screw employed for insertion revealed no issues.Consequently, the surgeon concluded that the device had deflected due to the patient's bone quality and shape, preventing insertion.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GAMMA4 - LAG SCREW D10.5X90MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
dannielle stap
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key24277549
Report Number0008031020-2026-00122
Device Sequence Number8365473
Product Code HSB
UDI-Device Identifier07613327510928
UDI-Public(01)07613327510928(11)250131(17)341231(10)K102E0C
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K213328
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number81600090S
Device Lot NumberK102E0C
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/08/2026
Supplement Date Received by Manufacturer02/11/2026
Initial Report FDA Received Date02/06/2026
Supplement Report FDA Received Date03/20/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/31/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-