| Catalog Number |
8065751511 |
| Medical Device Problem Codes |
Obstruction of Flow (2423); Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A customer reported that, during surgery foreign material was found stuck to irrigation/ aspiration (i/a) tip.Procedure details and surgery completion details are unknown.Information regarding patient harm was not reported.Additional information has been requested, none received till date.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in sections h.6 and h.11.A sample was not received at the manufacturing site for evaluation for the report of foreign material; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot number indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The investigation conducted a non-conformance review of the reported lot number.No deviations were identified and all corresponding production release specifications defined in the device master record were met.The photo evaluation confirms the presence of foreign material in the i/a tip port based on the evaluation of the information received; however, the investigation was unable to identify the root cause or origin of the reported event.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.All i/a tips are 100% visually inspected prior to being released from the manufacturing facility.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all manufacturer products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|