| Catalog Number |
88710100 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that outside of surgery, the unit doesn't work, it does not power on.During device evaluation it was discovered that the motor speeds were not stable.There was no patient involvement.Due diligence is complete; there is no additional information available.
|
| |
|
Additional Manufacturer Narrative
|
|
This event is recorded with zimmer biomet under (b)(4).This medwatch is being filed as an initial / final report based on information discovered during the device evaluation.Review of the most recent repair record determined the motor speeds were not stable and the reciprocating arm was worn.The motor and reciprocating arm were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.The root cause of the reported issue is attributed to component failure as the device exhibits wear and tear from repeated use.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.E1: telephone: (b)(6).G2: foreign - event occurred in netherlands.
|
| |
|