| Catalog Number |
165812 |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
01/06/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during use of foley catheter, customer was unable to empty the filled balloon, so they cut the catheter in half to empty it.
|
| |
|
Event or Problem Description
|
|
It was reported that during use of foley catheter, customer was unable to empty the filled balloon, so they cut the catheter in half to empty it.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed- other.Received a 2 way foley catheter and a straight tipped syringe.This sample will be tested for "difficult to remove".Inflated balloon with 10 ml solution using the returned syringe, noting strong resistance, and the catheter rested with no leaks noted.Attempted to deflate balloon passively and only 1 ml could be withdrawn.Using the in house luer lock syringe, deflated balloon in 3 minutes and 37 seconds with no cuffing noted.The complaint is against the return syringe.The labeling is found to be adequate.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.Correction: d,f,h.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|