A bipap avaps device was returned to a third-party service center with no initial alleged complaint.There was no report of serious or permanent harm or injury.There was no report of medical intervention.During the evaluation of the device at the third-party service center, the device was visually inspected, and evidence of foam particles was found inside the blower kit and blower foam degradation.Additionally, the service technician also noted dust contamination, fluid fail and the evaluation of the device data identified unrelated error codes.These error codes are consistent with normal device operation and expected use condition and are not considered reportable under medical device reporting requirements.No evidence was identified to suggest that these error codes contributed to or caused the reported event.The device was scrapped by the service center after the evaluation was completed.
|