Catalog Number 81570
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
In this event it is reported that a 30k fsi-sli-1000 insert,pkd was discovered upon opening the package that the inserts stem was bent at the end.The user tried using the insert and during use the nozzle become hotter than normal even with increased water flow.Reportedly, no injury.
 
Additional Manufacturer Narrative
Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device was not returned for evaluation.However, the lot number was provided and retained-product testing and/or dhr review are planned.The results will be submitted as they become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
30K FSI-SLI-1000 INSERT,PKD
Common Device Name
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer (Section G)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer Contact
dan eagar
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key24279139
Report Number2424472-2026-00513
Device Sequence Number9803293
Product Code ELC
UDI-Device IdentifierD003815701
UDI-PublicD003815701
Combination Product (Y/N)N
Initial Reporter CountryUP
PMA/510(K) Number
K052334/K970
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 05/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number81570
Device Lot Number00133755
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date01/30/2026
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/30/2026
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date02/06/2026
Supplement Report FDA Received Date05/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-