(b)(4).Date sent: 2/6/2026 d4: batch # 621d92 investigation summary the product was returned to ees for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the tlc75 device was received with no apparent damage and with a reload loaded in the device.The reload had the proximal 8 drivers up without staples, the remaining drivers down with staples present, and with the cartridge deck damaged.The returned reload had the swing tab in the locked position.The device was tested for functionality with a test reload and it fired, cut, and formed all staples as intended.The staple line and cut line were complete and the staples met the staple form release criteria.The damage to the reload is consistent with the device being clamped over a hard object.To mitigate the potential for staples getting into the reload and interfering with the knife during device firing, prior to reloading the device, clean in sterile solution and then wipe the anvil and reload channel to clean any formed but unused staples from the device.Additionally, proper care should be taken when placing the device on the tissue to be stapled, to ensure that no hard obstruction such as a clip is included with the tissue inside the jaws.The event reported was confirmed and is related to improper use of the device.The cartridge reload is designed to lockout as a safety feature if any staples have been fired from the cartridge reload.Failure to complete the stroke may result in incomplete staple line.Please reference the instruction for use for more information.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.Device history review a manufacturing record evaluation was performed for the finished device batch number 621d92, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|