• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO VILLALBA SPRINT QUATTRO SECURE S MRI SURESCAN; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO VILLALBA SPRINT QUATTRO SECURE S MRI SURESCAN; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number 693558
Medical Device Problem Codes Signal Artifact/Noise (1036); Fracture (1260); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Codes Shock from Patient Lead(s) (3162); Medical device site infection (4845)
Date of Event 01/16/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Continuation of d10: ddmb1d1 icd implanted (b)(6) 2020.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the implantable cardioverter defibrillator (icd) system was explanted and replaced due to vegetation on the right ventricular (rv) lead.The rv lead exhibited intermittent noise, non-physiologic/sensing integrity counters (sic) and a suspected lead fracture.The patient received an inappropriate shock and therapies were turned off.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the distal portion of the lead was returned and analyzed.The lv1/distal conductor was fractured in-vivo in the anchoring sleeve area.The distal conductors were fractured at 18.5 cm.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO SECURE S MRI SURESCAN
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key24280930
Report Number2649622-2026-03574
Device Sequence Number9862125
Product Code LWS
UDI-Device Identifier00613994446091
UDI-Public(01)00613994446091(17)140228
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/28/2014
Device Model Number693558
Device Catalogue Number693558
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/16/2026
Supplement Date Received by Manufacturer02/26/2026
Initial Report FDA Received Date02/06/2026
Supplement Report FDA Received Date03/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/09/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5076-45 LEAD.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention; Life Threatening;
Patient Age47 YR
Patient SexFemale
-
-