|
|
| Catalog Number |
MCP426H |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient underwent a gynecological procedure in 2026 and suture was used.It was reported by the sales representative that during a gynecological procedure, while performing skin closure, the tip of the needle broke off.An x-ray was performed on the patient to locate the broken piece of the needle.The x-ray was reviewed and cleared by the surgeon.Device is available for return.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested, and the following was obtained via: did the needle fall into the patient? if so, was the needle piece(s) retrieved during the same procedure? were x-rays taken to locate the needle piece(s)? what measures were implemented to retrieve the broken piece? was there any additional tissue damage resulting from the search for the needle piece? does a fragment of the needle remain in the patient¿s tissue? if so, is there any plan in place to remove the needle piece in the future? if so, could you please provide the scheduled date for the removal? in which tissue structure was the broken needle located? what is the surgeon¿s opinion of the potential consequences for the patient? could you kindly perform and document the follow-up attempt for the product return? furthermore, please ensure that the shipment tracking number is recorded in the notes or rmao sections.Please provide the source or name of person providing answers to follow-up questions: i do not have additional information available outside of the initial report.An x-ray was performed on the patient to locate the broken piece of the needle.The x-ray was reviewed and cleared by the surgeon.I do not know what the surgeon's definition of "cleared" implies.Surgeon is not available for follow up.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 3/9/2026.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: d9, h3, h4, h6.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: the product received for analysis was mcp426h.Visual inspection and metallurgical analysis were performed on the returned product.A failed suture needle was submitted to hsa for fractographic evaluation.The component was identified as product code mcp426h.It was requested that hsa assess the fracture mode of the failure.A fracture was observed at the tip of the needle.One side of the needle was received, the mating fracture surface was not provided for this evaluation.A scanning electron microscope was used to examine the fracture surfaces and surrounding area of the needle.The fracture surfaces were examined in multiple locations to determine the fracture mode.The evaluation revealed the fracture was composed of microvoid coalescence, which is evidence of a ductile fracture mode.Mechanical damage observed coincidental to the fracture provides additional evidence that the failure was induced by mechanical deformation leading to ductile overload.This was a ductile fracture.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
|
| |
|
Search Alerts/Recalls
|
|
|