| Catalog Number |
3910-900-093 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/08/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that the needle broke.All pieces were removed.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: the needle broke.The failure(s) identified in the investigation is consistent with the complaint record.The probable root causes could be: 1) use of excessive force, 2) attempt to pass through thick tissue or bone, 3) excessive tissue loaded into jaws, 4) attempt to load or pass incompatible suture, 5) technique error.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that the needle broke.All pieces were removed.
|
| |
|
Search Alerts/Recalls
|