W. L. GORE & ASSOCIATES, INC. GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Catalog Number TAC124515A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Post Operative Wound Infection (2446)
Date of Event 01/22/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As gore was unable to determine which device was involved in the event if any; additional device(s) implanted include: te4546a/(b)(6), udi: (b)(4).As gore was unable to determine which device was involved in the event if any; additional device(s) implanted include: te4546a/(b)(6), udi: (b)(4).W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may have not been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
The following details were reported to gore: on (b)(6) 2025, this patient underwent endovascular treatment of a zone 0 (ascending aorta) with bca (brachiocephalic artery) tevar treatment for aneurysm from a dissection.The patient was implanted with gore® tag® thoracic branch endoprosthesis device and gore® tag® conformable thoracic stent graft with active control system (tsb121706a, tgmr373115, tsb121706a, tac124515a, te4546a).The patient tolerated the procedure with good results.On (b)(6) 2026, the patient underwent reintervention to treat continued aneurysmal degeneration of the aneurysm due to infection of the patient¿s aorta and of the aneurysm.The side branch device was extended proximally and the tbe aortic component was extended with a ctag.The patient tolerated the procedure.The cause of the infection is not known.The physician reports that at the patient¿s 30-day follow-up computed tomography (ct) all looked good healthy and no device issues or concerns noted.The patient had observed device issues.The patient then stated they had a follow-up ct at another hospital in january and the aorta had gotten progressively worse.The physician does not know if there was undetected infection at the time of the index procedure, or the possibility of a sterility issue at the time of the index procedure.But root cause of the degeneration is thought to be infection of the aneurysm which has spread within the aorta and the grafts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL WOODY SPRINGS B/P
3450 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
laura crawford
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key24283686
Report Number2017233-2026-07124
Device Sequence Number20905372
Product Code MIH
UDI-Device Identifier00733132654253
UDI-Public(01)00733132654253(17)280517
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberTAC124515A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/22/2026
Initial Report FDA Received Date02/06/2026
Date Device Manufactured05/19/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age60 YR
Patient SexMale
Patient Weight82 KG
-
-