THORATEC SWITZERLAND GMBH CENTRIMAG 2ND GENERATION PRIMARY CONSOLE; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 201-90411
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/28/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the centrimag 2nd generation primary console failed during its yearly functional checkout.It did not meet iec 61000-4-5 standards.
 
Additional Manufacturer Narrative
Manufacturer's investigation conclusion: the reported event of the centrimag console not passing surge testing was confirmed via the console¿s functional analysis.The returned centrimag console (serial number (b)(6)) was received at the service depot, and the console did not withstand surge testing of up to 2.0 kv.The unit will be held by the service depot until parts become available for rework.The root cause for the reported event was determined to be related to the ac/dc power supply.A corrective action was initiated for this issue, and this centrimag console was manufactured prior to the implementation of this corrective action.Review of the device history record for the centrimag console, serial number (b)(6), showed the device was manufactured in accordance with manufacturing and qa specifications.The 2nd generation centrimag system operating manual provides information regarding emergency/troubleshooting in section 10.The recommended practice whenever there is a console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the back-up motor and console.Switch all components (console, motor, flow probe and cables) simultaneously to continue patient support, and then perform troubleshooting on the non-functioning system, when it is no longer being used for patient support.No further information was provided.The manufacturer is closing the file on this event.
 
Additional Manufacturer Narrative
No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIMAG 2ND GENERATION PRIMARY CONSOLE
Common Device Name
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
Manufacturer (Section G)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ   CH-8005
Manufacturer Contact
fredric gold
6035 stoneridge drive
pleasanton, CA 94588
7818528207
MDR Report Key24284059
Report Number3003306248-2026-00119
Device Sequence Number9962760
Product Code DWA
UDI-Device Identifier07640135140702
UDI-Public(01)07640135140702(11)211010
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K131179
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse
Remedial Action Notification
Type of Report Initial,Followup
Report Date (Section B) 02/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number201-90411
Device Catalogue Number201-90411
Device Lot Number8434145
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/28/2026
Supplement Date Received by Manufacturer02/05/2026
Initial Report FDA Received Date02/06/2026
Supplement Report FDA Received Date02/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/10/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
FA-Q224-HF-2
Patient Sequence Number1
Patient SexUnknown
-
-