| Catalog Number |
NOD12LPT |
| Medical Device Problem Codes |
Break (1069); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2026, a patient underwent an ultrasound-guided pcn percutaneous drainage catheter placement procedure using a navarre opti drain catheter.Post the procedure, the sure-lok tm thread had digression.The procedure was completed using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records will be performed.The device has not been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.Section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: the device history records have been reviewed, and this lot met all release criteria.Investigation summary: although the physical device was not returned for evaluation, three photos were provided for review.The three photos shows tangled thread on the catheter hub and product details were hidden behind catheter in first and second photo.Third photo shows product details which matched with tw details and tangled thread on the catheter hub.Based on the provided photo the reported thread break and separation issues cannot be confirmed for one device.The investigation is inconclusive for reported thread break and material separation issues for one device.Due to the absence of photos for remaining two devices investigation remains inconclusive.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.G3, h6 (method, result, conclusion), section a through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2026, a patient underwent an ultrasound guided pcn percutaneous drainage catheter placement procedure using a navarre opti drain catheter.Post the procedure, the sure loktm thread had digression.The procedure was completed using another device.There was no reported patient injury.
|
| |
|
Search Alerts/Recalls
|