Catalog Number 72201781
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Perforation (2001)
Date of Event 01/12/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a patient complication related to a product problem with a smith+nephew device.The reported issue(s) relate to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alter the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
It was reported that, during an acl procedure, while fixing the graft in the femur, the surgeon didn¿t want to mallet the tunnel by attaching the biosure ha 9mm x 25mm directly to the tunnel and with excessive force, damaging the tip of the screw and it broke.When removing the screw with the damaged tip, the surgeon removed the graft and cleaned it with the shaver blade.An additional bone hole was required.The procedure was resumed, without any delay, using an equivalent s+n back up.No further complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOSURE HA 9MM X 25MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key24290266
Report Number1219602-2026-00373
Device Sequence Number13300343
Product Code HWC
UDI-Device Identifier03596010611772
UDI-Public3596010611772
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K080358
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72201781
Device Lot Number51276192
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2026
Initial Report FDA Received Date02/09/2026
Date Device Manufactured12/18/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age36 YR
Patient SexMale
Patient Weight85 KG
-
-