• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION CORPORATION AVISTA? MRI; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION CORPORATION AVISTA? MRI; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number SC-2408-56
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/14/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Block b3: approximated based on the date the manufacturer became aware of the event.Additional suspect medical device components involved in the event: product family: scs-linear leads-mri.Upn: m365sc2408560, model: sc-2408-56, serial: (b)(6), batch: 7088187, udi: (b)(4).Product family: scs-lead fixation-mri: upn: m365sc43190, model: sc-4319, serial: na, batch: 36550425, udi: (b)(4).Product family: scs-lead fixation-mri: upn: m365sc43190, model: sc-4319, serial: na, batch: 33308888, udi: (b)(4).
 
Event or Problem Description
It was reported that the patients spinal cord stimulation (scs) leads had migrated.The patient underwent a revision procedure wherein the leads were explanted and replaced with a paddle lead.The explanted leads were kept by the facility.
 
Event or Problem Description
It was reported that the patients spinal cord stimulation (scs) leads had migrated.The patient underwent a revision procedure wherein the leads were explanted and replaced with a paddle lead.The explanted leads were kept by the facility.
 
Additional Manufacturer Narrative
Block b3: approximated based on the date the manufacturer became aware of the event.Additional suspect medical device components involved in the event: product family: scs-linear leads-mri.Upn: m365sc2408560.Model: sc-2408-56.Serial: (b)(6).Batch: (b)(6).Udi: (b)(4).Product family: scs-lead fixation-mri.Upn: m365sc43190.Model: sc-4319.Serial: na.Batch: (b)(6).Udi: (b)(4).Product family: scs-lead fixation-mri.Upn: m365sc43190.Model: sc-4319.Serial: na.Batch: 33308888.Udi: (b)(4).Sc-2408-56 sn: (b)(6).Investigation results: the device was not returned for analysis; therefore, a technical analysis could not be performed.Device history record: it was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review: review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Investigation conclusion: based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.Sc-2408-56 sn: (b)(6).Investigation results: the device was not returned for analysis; therefore, a technical analysis could not be performed.Device history record: it was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Investigation conclusion: based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVISTA? MRI
Common Device Name
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
erik sherburne
25155 rye canyon loop
valencia, CA 91355
7632920920
MDR Report Key24290767
Report Number3006630150-2026-00774
Device Sequence Number21394258
Product Code LGW
UDI-Device Identifier08714729904816
UDI-Public(01)08714729904816(17)260301(10)7083471
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P030017
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/01/2026
Device Model NumberSC-2408-56
Device Catalogue NumberSC-2408-56
Device Lot Number7083471
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/14/2026
Supplement Date Received by Manufacturer04/08/2026
Initial Report FDA Received Date02/09/2026
Supplement Report FDA Received Date05/01/2026
Date Device Manufactured03/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexMale
-
-