U.S. Department of Health and Human Services

MAUDE Adverse Event Report: UNOMEDICAL S.R.O. GENTLECATH GLIDE; CATHETER, URETHRAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

UNOMEDICAL S.R.O. GENTLECATH GLIDE; CATHETER, URETHRAL Back to Search Results
Model Number 421567
Medical Device Problem Code Sharp Edges (4013)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
End user states "he can feel a burr near the eyelet for his gc glide.".
 
Additional Manufacturer Narrative
Device 2 of 2.Based on the available information, this event is deemed to be a reportable malfunction.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Reporting site: 1049092.Manufacturing site: 3005778470.
 
Additional Manufacturer Narrative
Batch record review was conducted with the following findings: all catheters were produced on catheter machine a116 in march 2025.All raw materials used in the production of the affected lots were verified to be correct and within specification.No nonconformances were recorded during the production process of catheters.No similar complaint was received for lot 5c03363 and malfunction code uca-pmc07.02 catheter shaft, balloon, tip or eyelets too rough/jagged edges or too sharp.The affected rate is (b)(4) (2 affected pieces out of (b)(4)), which is within the (b)(4) aql limit applicable to this type of issue.The process parameters were adjusted within the validated window.The machine logbook was reviewed, and no records related to this issue were found.Additionally, no ncs were identified during the sterilization process.Based on the available documentation and investigation results, no manufacturing-related root cause was identified for this complaint.Should additional information become available, a follow-up report will be submitted.Fda registration number: reporting site: 1049092.Manufacturing site: 3005778470.
 
Event or Problem Description
To date no additional patient or event details have been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GENTLECATH GLIDE
Common Device Name
CATHETER, URETHRAL
Manufacturer (Section D)
UNOMEDICAL S.R.O.
priemyselny park 3,
michalovce 07101
LO  07101
Manufacturer (Section G)
UNOMEDICAL S.R.O.
priemyselny park 3,
michalovce 07101
LO   07101
Manufacturer Contact
pamela meadows
7815 national service road
suite 600
greensboro, NC 27409
3365424679
MDR Report Key24291802
Report Number3005778470-2026-00260
Device Sequence Number10166164
Product Code GBM
UDI-Device Identifier30768455143543
UDI-Public30768455143543
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K181206
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Distributor
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number421567
Device Lot Number5C03363
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/05/2026
Supplement Date Received by Manufacturer03/02/2026
Initial Report FDA Received Date02/09/2026
Supplement Report FDA Received Date03/02/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/18/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-