• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. POWERED PLUS COMPACT 45; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. POWERED PLUS COMPACT 45; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number PCEE45A
Medical Device Problem Code Excessive Heating (4030)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/26/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was preoperatively to a vats procedure; it was observed by the staff that the battery and handle appeared to be overheating and felt hot to touch.This was outside of the patient.Changed to a new one to complete the procedure.There was no patient consequence nor surgical delay reported.No additional information provided.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 2/9/2026.D4: batch#: 573d31.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one pcee45a device was returned with the anvil bent upwards and with a gst45g reload loaded on the device.The reload was received fully fired with damage on reload deck.Furthermore, the anvil knife slot was noted to be damaged.Upon analysis, the jaws and closure trigger were noted in the close position, the knife was noted to be slightly advanced.The knife reverse switch was activated and the knife returned to the home position as expected.After further evaluation, the analysis showed the knife wings were broken off.Only one wing of the knife was returned for analysis.No functional test was performed due to the condition of the device.It is possible that the device was clamped over an excess of tissue causing the anvil to bend and for the firing stroke and the staple form to be incomplete.It is possible that the combination of tissue and/or buttressing material compressed between the device jaws may have been beyond indicated thickness or tissue that cannot compress to the indicated range for the selected reload.This may have caused excessive force to be applied to the underside of the solid steel anvil leading to this damage.The damage to the reload is consistent with the device being clamped over a hard object.To mitigate the potential for staples getting into the reload and interfering with the knife path during device firing, prior to reloading the device, rinse the anvil and reload jaw in sterile solution and then wipe the anvil and reload jaw to clean any formed but unused staples from the device.Additionally, proper care should be taken when placing the device on the tissue to be stapled, to ensure that no hard obstruction such as a clip is included with the tissue inside the jaws.The damage to the knife is consistent with the device being clamped over an excess of tissue.If the firing continues the knife will remain inside the guide, and as the knife is attempting to pull the anvil towards the reload to form the staples it will result in the damage observed on the knife.The event described of device handle temperature and battery pack temperature could not be confirmed as no temperature problems were detected.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number: 573d31, and no non-conformance's were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERED PLUS COMPACT 45
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 00969
*   00969
3035526892
MDR Report Key24293083
Report Number3005075853-2026-01151
Device Sequence Number9415245
Product Code GDW
UDI-Device Identifier10705036015338
UDI-Public(01)10705036015338(10)573D31(17)280430
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K163454
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberPCEE45A
Device Lot Number573D31
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/02/2026
Initial Report FDA Received Date02/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/08/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-