| Model Number |
ENF-GP2 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3 - the date of event is unknown.A supplemental report will be submitted if provided.The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported the rhino-laryngo fiberscope was dropped and found to have fiberglass inside of the device.There were no reports of patient harm.
|
| |
|
Event or Problem Description
|
|
No additional information received from customer.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the final investigation.The device was returned to olympus for inspection, and the reported failure was confirmed.A definitive root cause could not be identified.Based on the results of the investigation, the most probable cause of the complaint is cause traced to component failure indicating expected or random component failure without any design or manufacturing issue.Olympus will continue to monitor field performance for this device.
|
| |
|