A sample was not received at the manufacturing site for evaluation for the report; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The photos attached to the complaint was reviewed by the manufacturing site.The photo 1 and 2 is of a pak label, the reported product and lot information is confirmed.Reported issue cannot be confirmed from photos.The investigation conducted a nonconformance review of the reported lot.Two nonconformances were found related to broken spacer and spacer short shot.This non-conformance could have potentially contributed to the reported event.Based on the evaluation of the information received, the investigation was unable to identify the root cause or origin of the reported event.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.However, two non-conformance records have been initiated in relation to the related non-conformances identified within the non-conformance review.A nonconformance investigation has been completed in relation to the trend identified for probes with would not cut or actuate issues.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
|