SYNTHES GMBH OSCILLATING SAW ATCH, LARGE, WITH KEY; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
| Catalog Number |
532.026 |
| Medical Device Problem Codes |
Complete Loss of Power (4015); Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11.Additional narrative a visual and functional assessment was performed on the device.The following steps failed: general condition.Check oscillation frequency with frequency meter.During the service evaluation the following failures were identified: device interaction (2+ devices): unintended/unexpected disengagement.Visual: sharp edges.Functional: does not function.Conclusion: failure reported is confirmed.Root cause: improper maintenance.
|
| |
|
Event or Problem Description
|
|
It was reported that during service and evaluation, it was determined that the saw attachment device had unintended/unexpected disengagement.It was noted in the service order that the device was not working properly.There was no human patient involvement as this device is intended for veterinary purposes.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2026.All available information has been disclosed.
|
| |
|
Search Alerts/Recalls
|
|
|