Based on the information provided ¿ which may include the returned device (if available), photographs, videos, event description, and any additional field input ¿ arthrex has determined the most likely cause.Neither the method used to prepare the bone nor the bone quality encountered during the procedure were provided.The most likely attributed to use-related factors such as improper bone preparation, misaligned insertion, or incomplete bone preparation at the insertion site.Per dfu-0054-eo revision 7, c.Contraindications, 7: the use of this device may not be suitable for patients with insufficient or immature bone.The physician should carefully assess bone quality before performing orthopedic surgery on patients who are skeletally immature.E.Warnings: 6: preoperative and operating procedures, including knowledge of surgical techniques and proper selection and placement of the device, are important considerations in the successful utilization of this device.The appropriate arthrex delivery system is required for proper implantation of the device.7.Avoid excessive impaction during anchor insertion as this could lead to inserter damage and/or breakage.If insertion resistance is encountered, do not impact harder.Replace the implant and repeat the drilling/insertion process.G.Precautions, 3: fastak suture anchors only: stopping and restarting the drill may result in excess torque being applied to the implant, which may cause the device to fail.
|