• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BERLIN HEART GMBH EXCOR®; VENTRICULAR (ASSIST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BERLIN HEART GMBH EXCOR®; VENTRICULAR (ASSIST) BYPASS Back to Search Results
Model Number P50P-001
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Brain Injury (2219)
Date of Event 01/06/2026
Type of Reportable Event Death
Additional Manufacturer Narrative
The excor blood pump pu valves; 50 ml in/out; ø 12 mm, lvad, s/n (b)(6) and rvad, s/n (b)(6), are used on the patient from (b)(6) 2026 until the pumps were removed from the patient on (b)(6) 2026.The event occurred on 2026-01-06, which is (5 days) after the pumps were placed on the patient.The patient was being supported with an excor ikus driving unit.We have reviewed the production records of the excor blood pumps.The pumps were produced according to our specifications.The excor rvad blood pump s/n (b)(6) information is as follows: the model number for this lot: p50p-001 the manufacture date for this lot: 2025-08-12 the expiration date for this lot: 2028-07-19 udi: (b)(4).
 
Event or Problem Description
The site contacted berlin heart clinical affairs (ca) on (b)(6) 2026 to report that the patient presented on (b)(6) 2026 with an acute change to neurologic exam with bilateral fixed and dilated pupils with npi is 0.Two head ct were performed showing diffuse loss of gray-white matter differentiation through the bilateral cerebral and cerebellar hemispheres, slightly more conspicuous compared to prior exam.Redemonstration of diffuse effacement of the extra-axial spaces and basal cisterns.New mild cerebellar tonsillar herniation.The ventricles are slit like, slightly decreased compared to prior study.There is no significant midline shift.There is no mass or intracranial hemorrhage.The osseous structures appear normal.Scattered paranasal sinus opacification predominantly in the sphenoid and ethmoid sinuses.Opacification of bilateral mastoid air cells and middle ear cleft, left greater than right.The patient continued to be monitored, and the patient was normothermic, normotensive, without paralysis/sedation and without significant metabolic derangements at that time.On (b)(6) 2026 the patient was withdrawn from support and expired due to poor outcome prognosis.The patient was being supported with the excor system in bi-vad configuration.The berlin excor blood pump functioned as intended maintaining full fill and eject.No deposits were noted in both the pumps at the time of the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXCOR®
Common Device Name
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
BERLIN HEART GMBH
wiesenweg 10
berlin, 12247
GM  12247
MDR Report Key24299758
Report Number3008454189-2026-00001
Device Sequence Number10153526
Product Code DSQ
UDI-Device Identifier04260090040010
UDI-Public(01)04260090040010(17)270729
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/09/2026
Report Date (Section F) 02/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberP50P-001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date01/12/2026
Device Age17 MO
Event Location Hospital
Date Report to Manufacturer01/12/2026
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date02/09/2026
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age5 YR
Patient SexMale
Patient Weight32 KG
Date Report Sent to FDA02/09/2026
-
-