| Model Number |
693558 |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Fracture (1260); High impedance (1291); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Code |
Shock from Patient Lead(s) (3162)
|
| Date of Event |
01/16/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the right ventricular (rv) lead was suspected to be fractured.Subsequently, the patient received an inappropriate shock.Additionally, the lead triggered alerts for noise and undefined high pacing impedance.The lead triggered a lead integrity alert (lia) for high rate non-sustained episodes and sensing integrity counter (sic).The lead was explanted and replaced.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|