SIEMENS HEALTHCARE DIAGNOSTICS INC. ATELLICA IM TOXOPLASMA IGG (TOXO G); ENZYME LINKED IMMUNOABSORBENT ASSAY, TOXOPLASMA GONDII Back to Search Results
Model Number N/A
Medical Device Problem Code False Positive Result (1227)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/11/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This incident occurred outside the united states (ous).The ous customer reports an observation of discordant toxoplasma igg (toxo g) results for a pregnant patient with reagent lot 298.The initial atellica im toxo g result (sample # (b)(6) on (b)(6) 2025) was reactive (positive).The result was reported to the physician and questioned.The sample was sent to other labs for alternate method testing.Results from the alternate methods were non-reactive (negative).A second sample from the same patient was drawn on a different date (sample # (b)(6) on (b)(6) 2026) and resulted positive with atellica im toxo g.The sample was sent to other labs for alternate method testing.Results from the alternate methods were non-reactive (negative).There are no allegations of patient or user intervention or adverse health consequences due to the event.Siemens is investigating.Note: this report is for reporting the discordant result obtained on (b)(6) 2025.A separate report is submitted for the discordant result obtained on (b)(6) 2026.
 
Event or Problem Description
The customer reports discordant toxoplasma igg (toxo g) results for a pregnant patient with reagent lot 298.The initial atellica im toxo g result (sample # (b)(6) on (b)(6) 2025) was reactive (positive).The result was reported to the physician and questioned.The sample was sent to other labs for alternate method testing.Results from the alternate methods were non-reactive (negative).A second sample from the same patient was drawn on a different date (sample # (b)(6) on (b)(6) 2026) and resulted positive with atellica im toxo g.The sample was sent to other labs for alternate method testing.Results from the alternate methods were non-reactive (negative).There are no allegations of patient or user intervention or adverse health consequences due to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATELLICA IM TOXOPLASMA IGG (TOXO G)
Common Device Name
ENZYME LINKED IMMUNOABSORBENT ASSAY, TOXOPLASMA GONDII
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict ave
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
333 coney st.
registration # 1219913
east walpole MA 02032
Manufacturer Contact
stacy loukos
333 coney st.
east walpole, MA 02032
MDR Report Key24301362
Report Number1219913-2026-00033
Device Sequence Number5240277
Product Code LGD
UDI-Device Identifier00630414600123
UDI-Public(01)00630414600123(10)53871298(17)260616
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K012183
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/16/2026
Device Model NumberN/A
Device Catalogue Number10995699
Device Lot Number53871298
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/27/2026
Supplement Date Received by Manufacturer03/24/2026
Initial Report FDA Received Date02/10/2026
Supplement Report FDA Received Date06/09/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/16/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age28 YR
Patient SexFemale
-
-