• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET SAS VOLISTA ACCESS; LIGHT, SURGICAL, CEILING MOUNTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS VOLISTA ACCESS; LIGHT, SURGICAL, CEILING MOUNTED Back to Search Results
Model Number ARDVCS209004A
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The initial reporter was clinical engineer.Event site name: (b)(6).Getinge became aware of an issue with one of surgical lights - volista access.The designated complaint unit employee confirmed based on photographic evidence that paint was chipping on fork.We decided to report the issue in abundance of caution as any particles falling off into sterile field or during procedure may cause contamination.Based on the information collected, it was established that when the event occurred, the surgical lights did not meet its specification, since paint chipping could be considered as technical deficiencies, and in this way the device contributed to event.It is unknown if claimed devices were or were not being used for patient treatment or diagnosis when the event took place.A root cause analysis for paint peeling problem was performed by subject matter experts and concluded that all maquet sas products comply with: iec 60601-1 ed.2.0 & ed.3.0 general requirements for basic safety and essential performance.Iec 60601-2-41 particular requirements for the safety of surgical luminaires and luminaire for diagnosis.Paint definition pfc066.This procedure defines maquet sas¿s requirements for all painted parts.Disinfection products test: the aim of these tests is to detect any incompatibility with disinfectant.The paint chip or paint damages are due to: impacts, collisions (abnormal use).The operating manual includes the instructions to pre-position the arms prior to use, in order to prevent damages.To prevent any similar incident, it is recommended to avoid collisions between devices.Visual inspections during the cleaning allow to detect the painting defect, we recommend to perform corrective maintenance to rectify the default after its detection.Minor paint chip can be repaired with touch up paint, nevertheless, the parts impacted by serious damage must be replaced.Getinge shall continue to monitor for any further events of this nature and does not propose any further action at this time.
 
Event or Problem Description
Getinge became aware of an issue with one of surgical lights - volista access.The designated complaint unit employee confirmed based on photographic evidence that paint was chipping on fork.We decided to report the issue in abundance of caution as any particles falling off into sterile field or during procedure may cause contamination.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOLISTA ACCESS
Common Device Name
LIGHT, SURGICAL, CEILING MOUNTED
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer (Section G)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer Contact
sylviane boiron bourgeois
parc de limere
avenue de la pomme de pin
ardon 
MDR Report Key24302112
Report Number9710055-2026-0000026
Device Sequence Number5315589
Product Code FSY
UDI-Device Identifier03700712401320
UDI-Public(01)03700712401320
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K130513
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberARDVCS209004A
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/30/2026
Initial Report FDA Received Date02/10/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/17/2020
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-