(b)(4) date sent: 2/10/2026 d4: batch #482c66.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one ecr60w reload was received unfired, with the pan partially dislodged from the left proximal side and an open package.No functional test was performed due to the condition of the reload.No conclusion could be reached on what may have caused the reload pan to dislodge.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device ecr60w with lot number 519c13; and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device batch number 482c66, and no non-conformances related to the reported complaint condition were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|