| Model Number |
LUCAS |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Stryker evaluated the customer's device and was able to duplicate the reported issue.Investigation found the compression module did not operate smoothly and the communication cable had a cut.Stryker replaced the device's compression module, communication cable, protective board and hood to resolve the reported issue.After completing other unrelated repairs, proper device operation was observed through functional and performance testing.The device was returned to the customer for use.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device alarmed and paused.This state the device would not be able to perform automated chest compressions.If a device malfunctions, the device operating instructions indicate that the user is to remove the device and immediately start manual chest compressions.There was no patient involvement reported with the event.
|
| |
|
Search Alerts/Recalls
|