| Model Number |
GIF-XP170N |
| Medical Device Problem Codes |
No Display/Image (1183); Display or Visual Feedback Problem (1184); Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: the white screen with no image was due to component failure; however, the cause of the dirt on the objective lens could not be established.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the gastrointestinal videoscope exhibited dirt on objective lens and showed a white screen with no image.There was no patient involvement.
|
| |
|