| Catalog Number |
ECH45L |
| Medical Device Problem Codes |
Failure to Cut (2587); Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 2/10/2026.D4: batch#: unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot/batch number: a98h8g, and no non-conformance's were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unknown procedure, it was observed that the cutting function on the device did not work after the use of the green reload; and that the stapler simply clamped.Then after they rinsed and cleaned the device according to the instructions, they used another reload, but now the stapler would not close properly and there was a gap between the jaws, rendering the device unusable.There was no patient consequence nor surgical delay reported.No additional information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent 2/24/2026.The device upon which this medwatch is based has been received, however, the evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 3/19/2026.D4: batch #547d93.Corrected data: d4 (udi, expiration date), h4.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one ech45l device was returned with no apparent damage, with a tyvek, and with no reload present.The device was tested for functionality in the straight position with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples met the staple release criteria.The event described could not be confirmed as the device performed as intended and it opened and closed without any difficulties noted.The event log was reviewed.An event was found that typically indicates that the knife encountered the firing safety lockout mechanism.This mechanism ensures that the knife cannot deploy without an unfired reload installed.It is possible that it occurred due to improper installation of the reload.Always ensure that the retaining cap is in place when a reload is installed into a device.See the ifu for more details.In addition, the log indicates that articulation was attempted while the device was closed.Articulation function is blocked when the device is closed to prevent inadvertent patient injury.When the home button is pressed while the jaws are closed the device will attempt to retract the knife.When any of the other articulation buttons are pressed while the jaws are closed the device will provide haptic feedback indicating that articulation function is blocked.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 547d93, and no non-conformances were identified.
|
| |
|