MAUDE Adverse Event Report: ANGIODYNAMICS, INC. SOLERO; MICROWAVE TISSUE ABLATION APPLICATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS, INC. SOLERO; MICROWAVE TISSUE ABLATION APPLICATOR Back to Search Results
Model Number 700106002-US
Medical Device Problem Codes Fracture (1260); Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Foreign Body In Patient (2687); Device Embedded In Tissue or Plaque (3165)
Date of Event 01/06/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is not available to be returned to the manufacturer for evaluation.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Event or Problem Description
A distributor reported an end user experienced an issue when using a 19cm solero applicator.The applicator was tested prior to percutaneous insertion for 10sec @ 60 watts.The unit was set for 3min.@ 100 watts for the procedure.The applicator was placed in a solitary liver lesion, segment vi, when a high reflected power message was received.No ablation had been performed.The message was unable to be cleared, therefore the applicator was removed and inspected, when it was observed the tip had detached and was retained inside the patient's liver.The applicator was replaced with another and the procedure was completed.
 
Event or Problem Description
Additional information received: it was later reported that the patient is doing fine with no adverse impact.A ct scan had been performed to look for any possible complications, in which none were found.The tip of the applicator had been planned for surgical removal however it could not be done due to the patient having cirrhosis and not fit for surgery.The patient has resumed his normal activities and continues to be on close follow up.
 
Additional Manufacturer Narrative
The customer's reported complaint description of probe ceramic tip was fractured and detached was confirmed via picture provided, however, no probe complaint sample was returned.Without receiving product for evaluation a definitive root cause for this event cannot be determined.Potential root cause is a thermal event during the ablation cycle.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.The review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: the directions for use (dfu) which is supplied with this device contains the following statements: intended use/indications for use: the solero microwave tissue ablation (mta) system and accessories are indicated for the ablation of soft tissue: during open, laparoscopic, or percutaneous procedures.Testing the device: prior to placing the device in the target tissue for ablation, place the tip of the device (including the black shaded zone of the shaft) in a container of sterile water or saline and activate the device at 100 watts for 10 seconds to ensure that the system is functional.Warning: do not initiate the procedure/anesthesia until the applicator has been connected, primed, and the generator status bar indicates "ready".Precautions: avoid placing lateral forces on the applicator tip during placement or removal.Always use the lowest power and shortest time necessary to achieve the targeted ablation.Each applicator may be used to ablate up to three separate areas of target tissue for each patient at the maximum power and time setting.Inspect the applicator after each ablation.If the applicator appears damaged, utilize another applicator for subsequent ablations.Warning: do not bend the applicator as this may impair the function of the cooling system and may damage the microwave feed line inside the applicator.Warning: when using percutaneously the tip of the antenna may be used to puncture the skin at the point of insertion.Use the minimum force necessary to achieve this and take care not to over advance the applicator.Refer to the shaft depth markings to monitor placement depth.Take care to not bend the tip as it may cause damage to the device.Operational instructions: inspect all devices and packaging for damage prior to use.Do not use any devices that are damaged or if the sterile barrier is breached.Do not use the solero generator if it has been dropped or damaged.Warning: after each ablation inspect the applicator for any damage.If any damage is observed the applicator should be discarded and replaced with a new applicator.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOLERO
Common Device Name
MICROWAVE TISSUE ABLATION APPLICATOR
Manufacturer (Section D)
ANGIODYNAMICS, INC.
603 queensbury avenue
queensbury NY 12804
Manufacturer (Section G)
ANGIODYNAMICS, INC.
603 queensbury avenue
queensbury NY 12804
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
MDR Report Key24308125
Report Number1319211-2026-00026
Device Sequence Number15868195
Product Code NEY
UDI-Device IdentifierH787700106002US0
UDI-PublicH787700106002US0
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K162449
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number700106002-US
Device Lot Number5852274
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/22/2026
Supplement Date Received by Manufacturer02/11/2026
Initial Report FDA Received Date02/10/2026
Supplement Report FDA Received Date03/09/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/05/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age72 YR
Patient SexMale
-
-