MENTOR TEXAS MENTOR SMOOTH ROUND MODERATE PLUS PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Catalog Number |
3502300 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
01/17/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a primary breast augmentation surgery with implantation of a 300cc mentor smooth round moderate plus profile saline breast implant prosthesis.Post-operatively, the patient was diagnosed with a deflated right implant during a physical exam.As a result, the patient underwent bilateral exchange with mentor saline implants on (b)(6) , 2026.
|
| |
|
Additional Manufacturer Narrative
|
|
On february 18, 2026, the mentor analysis lab received the suspect medical device for evaluation.Paperwork received with the device indicated the device was also damage during explantation.On february 25, 2026, mentor completed an evaluation on the returned device.Device evaluation: mentor then conducted visual inspection and microscopic examination of the device.Visual analysis of the returned device determined that the breast implant was found to have three (3) tears and two (2) missing material zones all of them in the posterior view.Tear a, b & c measuring approximately 2.8 cm, 1.4 cm & 1.6 cm respectively.The missing material zones a & b measuring approximately 0.6 cm x 0.4 cm and 0.2 cm x 0.3 cm respectively.A microscopic examination was performed on the edges of the tear a, and parallel striations were found in an area of the tear on the posterior view, measuring approximately 0.1 cm.Parallel striations are consistent with markings made by a sharp object perforating the implant shell.The cause in the remaining area of the tear a could not be identified.Regarding the remaining tears (b & c) and missing material zones (a & b), a microscopic examination was performed, and the cause of the tears could not be identified.Therefore a thickness measurement was conducted on the shell at the tears, and the thickness was within manufacturing specifications for each of the tears and missing material zones.Although no conclusion could be reached on the cause of the tears (b & c) and missing material zones (a & b), the instructions for use contain the following caution: causes of deflation of saline implants include but are not limited to the following events: crease fold failure, intraoperative or postoperative trauma, excessive stresses or manipulations as may occur during normal, daily routines including customary and purposeful trauma such as that which can occur during vigorous exercise, athletics, etc.Manufacturer¿s reference number: ((b)(4).
|
| |
|