MAUDE Adverse Event Report: COVIDIEN AUTO SUTURE; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN AUTO SUTURE; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number OMS-T10SB
Medical Device Problem Codes Detachment of Device or Device Component (2907); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/29/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that during a laparoscopic hernia procedure, the grey valve of the balloon trocar dislodged and dropped into the abdominal cavity during mesh insertion, and the valve seal was damaged.The grey valve was retrieved using laparoscopic instruments to resolve the issue and complete the case.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Additional information: g3, h2, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the the returned product noted that the main valve seal was disengaged.The balloon, lock collar, and converter doors were intact.The inflation syringe was not received.The obturator was not received.Functional testing found that the balloon was inflated using a test inflation syringe, it was properly formed and no leaks detected.The lock collar moved up and down the cannula and snapped securely in place.It was reported that the seal was damaged and disengaged.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.Internal process improvements have been initiated to mitigate this issue.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTO SUTURE
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
COVIDIEN
sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
COVIDIEN
sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key24309039
Report Number2647580-2026-00320
Device Sequence Number9678706
Product Code GCJ
UDI-Device Identifier10884521081086
UDI-Public(01)10884521081086(17)280731(10)P5H0952(20)02
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K924011
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOMS-T10SB
Device Catalogue NumberOMS-T10SB
Device Lot NumberP5H0952
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/29/2026
Supplement Date Received by Manufacturer04/17/2026
Initial Report FDA Received Date02/10/2026
Supplement Report FDA Received Date04/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/15/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-