• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: UNOMEDICAL DEVICES S.A. DE C.V. INSET II; UNO INSET II 60/9 PINK TCAP 10PK INT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

UNOMEDICAL DEVICES S.A. DE C.V. INSET II; UNO INSET II 60/9 PINK TCAP 10PK INT Back to Search Results
Model Number 1002822
Medical Device Problem Code Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 01/21/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial and final mdr (b)(4) device 3 of 3.
 
Event or Problem Description
Reference number (b)(4) event occurred in canada.It was reported that patient faced infusion set leaking issue on (b)(6) 2025 and the infusion set tubing was split in half due to which the patient faced hyperglycemia event.The issue occured with 3 infusion sets and was occured in sometime in (b)(6) 2025, event 2: end of august 2025, event 3: (b)(6) 2026.The blood glucose level was high and the patient was treated with correction bolus via pump.The patient replaced the infusion set and resumed insulin deliveries successfully.No further information available.
 
Additional Manufacturer Narrative
Supplemental report 01 (b)(4) mdr 3003442380-2026-00505.Correction: this mdr is being submitted to correct the submitted common device name under d2, model number, serial number, primary unique device identifier (udi) number under d4.Additional information - this mdr is being submitted to include the below: h6: investigation results under type of investigation, investigation findings, investigation conclusions, h11: investigation summary.Complaint investigation results: electronic quality management system (eqms) search: a query was run in the eqms on 05-mar-2026 against "lot number" "6014835" and similar malfunction codes: tubing damage - significant / extensive, tubing is split (along the length of the tubing) or has pinholes.The review confirmed that lot 6014835 and the identified failure mode are not associated with any nonconforming reports (ncr)s or corrective and preventive action (capa)s of the same or similar nature.Similar complaints search: a query was run in the eqms on 05-mar-2026 against "lot number" criteria equal "6014835" and similar malfunction codes: tubing damage - significant / extensive, tubing is split (along the length of the tubing) or has pinholes.The count of complaints is 1.The complaints numbers are (b)(4).Device history record (dhr) review: the lot 6014835 was manufactured according to the work instruction (wi) version 125 and manufactured in the line inset 1, on 14-aug-2025 with a total of (b)(4) units.Glue-tubing lot: the lot 5h00090 was manufactured according to wi version 69 and manufactured in the machine sc01, on 07-aug-2025 with a total of (b)(4) units.The lot 5h00101 was manufactured according to wi version 69 and manufactured in the machine sc01, on 08-aug-2025 with a total of (b)(4) units.The device history record (dhr) was reviewed in accordance with applicable procedures.All required in-process and final tests were completed and met specified requirements.No deviations were identified, and no maintenance events were recorded that relate to the complaint code.Conclusion: dhr review supports compliance with manufacturing and quality requirements; no issues noted.Visual evidence review: no photo was provided to support visual confirmation of the reported issue.Conclusion: unable to perform visual verification; assessment based on available documentation only.Capa determination assessment - conclusion summary of complaint investigation: based on the investigation, no further investigation is required.The record will be closed and monitored through tracking and trending per wi (monthly trips and alerts).Complaint unconfirmed: following investigation the reported failure could not be confirmed for this complaint.Conclusion summary of complaint investigation: as a result of the following a comprehensive review was conducted, including eqms queries, similar complaint searches, device history record review, visual evidence assessment, and capa determination.No ncrs or capas of the same or similar nature were found for lot 6014835 and related malfunction codes for leakage.One complaint was identified for this lot, but no trend or systemic issue was detected.The manufacturing records confirmed that the lot was produced in compliance with all requirements, with no deviations or maintenance events noted.No photo evidence was provided, so the assessment was based on documentation only.Based on these results, no manufacturing or quality issues were identified, and no further investigation is required.The record will be closed and monitored through routine tracking and trending.
 
Event or Problem Description
To date no additional patient or event details have been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSET II
Common Device Name
UNO INSET II 60/9 PINK TCAP 10PK INT
Manufacturer (Section D)
UNOMEDICAL DEVICES S.A. DE C.V.
ave. fomento industrial l9 m3
parque industrial del norte
reynosa, tamaulipas, c.p., 88736
MX  88736
Manufacturer Contact
aaholmvej 1-3
osted
lejre, 4320
MDR Report Key24311709
Report Number3003442380-2026-00505
Device Sequence Number12155862
Product Code FPA
UDI-Device Identifier05705244018273
UDI-Public05705244018273
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K032854
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1002822
Device Lot Number6014835
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/27/2026
Supplement Date Received by Manufacturer03/05/2026
Initial Report FDA Received Date02/11/2026
Supplement Report FDA Received Date03/16/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/14/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age45 YR
Patient SexFemale
-
-